Abstract

A flow method of high-performance liquid chromatography (HPLC) seperation and chemiluminescence (CL) detection for sensitive vardenafil analysis in dietary supplements was developed. The vardenafil separation was achieved on a C18 column at 30°C using ethanol-H3PO4 and ethylenediaminetetraacetic acid disodium salt (Na2EDTA) aqueous solution (25 : 75, v/v%) as mobile phase. The followed continuous CL detection was conducted based on the strong CL enhancement by the presence of vardenafil to luminol-K3Fe(CN)6 reaction in alkaline medium. At the flow rate of 0.8 mL/min, the vardenafil retention time (tR) was 6.4 min. Factors that affected the HPLC resolution and CL detection were studied and optimized. The calibration curve obtained for vardenafil standard was linear in concentration range of 8.0 × 10−7 ~ 1.0 × 10−4 mol/L. The relative standard deviations (RSD) of intraday and interday precision were less than 3.5%. The proposed method was applied to the vardenafil determination in oral liquid, wine, and capsule samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.