Abstract

A high performance liquid chromatographic (HPLC) method with chemiluminescence (CL) detection was developed for the sensitive determination of 1, 3-diaminopyrene (1, 3-DAP), 1, 6-DAP, 1, 8-DAP and 1-aminopyrene (1-AP). The HPLC conditions were as follows: column, Cosmosil 5C18 (4.6mm i.d.×250mm); mobile phase, 10mM imidazole- perchloric acid buffer(pH 7.6)-acetonitrile (1:1, v/v); CL reagent, 0.02mM bis(2, 4, 6-trichlorophenyl)oxalate (TCPO) and 15mM hydrogen peroxide in acetonitrile. The oxidative degradation of DAPS and AP in the presence of metals was prevented by adding ascorbic acid to sample solutions. The calibration curves were straight over 2 orders of magnitude for all analytes, and their detection limits (as S/N was 3) were in the sub-fmol range. Dinitro- and nitropyrenes in sooty emissions of diesel- and gasoline-engine cars could be determined by this HPLC after reductive conversion into DAPs and AP, respectively, by refluxing the samples in the presence of sodium hydrosulfide.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.