Abstract

A new harmonic coil measurement system based on a dynamic signal acquisition device has been successfully developed to check the field quality of the quadrupole magnet for the CSNS/RCS, which operates at the 25Hz excitation cycle with a DC bias. It was designed to acquire multiple channels of data with a wide dynamic range of input signals, which are typically generated by a harmonic coil and an encoder. A dedicated algorithm was developed in LabView code to identify over specified intervals, synchronized to the coil’s rotation in the magnetic field. Through full integration of hardware and software, the traditional device (PDI 5025) is replaced successfully. This paper summarizes the characteristics of the system and presents the results of DC measurements.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call