Abstract

Reconstruction in electrical impedance tomography (EIT) is a nonlinear, ill-posed inverse problem. Based on point-shaped training and evaluation data, the "Graz consensus Reconstruction algorithm for EIT" (GREIT) constitutes a universal, homogenous method. While this is a very reasonable approach to the general problem, we ask the question if an optimized reconstruction method for a specific application of EIT, i.e. thoracic imaging, can be found. Instead of point-shaped training data we propose to use spatially extended training data consisting of eigenimages. To evaluate the quality of reconstruction of the proposed approach, figures of merit (FOMs) derived from the ones used in GREIT are developed. For the application of thoracic imaging, lung-shapes were segmented from a publicly available CT-database (www.dir-lab.com) and used to calculate the novel FOMs. With those, the general feasibility of using eigenimages is demonstrated and compared to the standard approach. In addition, it is shown that by using different sets of training data, the creation of an individually optimized linear method of reconstruction is possible.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.