Abstract
A gradient high-performance liquid chromatographic (HPLC) test procedure is developed and evaluated for its ability to establish the levels of impurities and remaining synthetic precursors in 2-[4-(1-hydroxy-4-[4-(hydroxydiphenylmethyl)-1-piperidinyl]-butyl)-phenyl]-2-methylpropionic acid. A gradient program with a mobile phase of 0.02 M sodium phosphate buffer and 0.004 M sodium perchlorate in acetonitrile-water (approximately pH 2.5) is used with a Spherisorb C6 column. The acetonitrile composition is increased linearly from 40% to 65% over a 45-min period and held at 65% for 20 min. UV detection at 210 nm is used to quantitate all components. The procedure is validated for accuracy using spiked levels (0.1% to 1.5%, w/w) with two suspected impurities, the synthetic precursors. A multiday repeatability study using two different Spherisorb C6 columns and HPLC systems shows consistent impurity quantitation results with one production lot of the bulk compound.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.