Abstract

Among quantitative lateral flow immunoassays (LFIAs) and assay readers, a system utilizing a magnetic field sensor for a reader and magnetic nanoparticle based-LFIA has recently been receiving a lot of attention because of its promising advantages such as high sensitivity and specificity. In this research, the quantitative LFIA reader platform utilizing the giant magnetoresistive (GMR) sensor was designed. The permanent magnets standing diagonally opposite to each other magnetized magnetic label particles and bias the sensor simultaneously to maximize the quantification capability. The performance of the reader prototype was evaluated with the imitative LFIA, which superparamagnetic nanoparticles were dispensed on the porous membrane. The system output changes due to the magnetic particle concentration change of 1mmol/L was 0.67μT, and 0.37mV of sensor output decreasing was observed when the temperature increased 1° in Celsius.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.