Abstract

A generic sample preparation method for the determination of acrylamide in foods was developed. The method entails extraction with methanol, purification with Carrez I and II solutions, evaporation and solvent change to water, and cleanup with Oasis HLB solid-phase extraction (SPE) cartridge. The final extract was analyzed by liquid chromatography-mass spectrometry (LC–MS) for quantitation. The chromatographic separation was performed on ODS-3 column using the isocratic mixture of 0.01 mM acetic acid in 0.2% aqueous solution of formic acid at a flow rate of 0.6 ml/min at 25 °C. The recoveries of acrylamide from potato chips, biscuits and coffee ranged between 92.8 and 101.5% with relative standard deviations of 4.1% or less. The limit of detection (LOD) and the limit of quantitation (LOQ) were 2 ng/g and 6 ng/g in the basis of signal to noise ratios of 3:1 and 9:1, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.