Abstract

Reconstruction of primary-photon energy spectra from pulse-height distributions obtained in a Compton spectrometer has earlier been performed under the assumption that coherent scattering in the scatterer is negligible. This holds for most clinical X-ray units operated in the range 40-150 kV-. In mammography, and to some extent in dental radiography, the relatively high frequency of low-energy photons (<30 keV) in the primary beam makes it necessary to extend the algorithms to allow for significant contribution of coherent scattering. This extension is performed as a perturbation calculation to the algorithms developed earlier in which a modified Klein-Nishina scattering cross section was taken as the total scattering cross section. Comparison with energy spectra measured in the primary beam indicates that the Compton spectrometer with the extended algorithm is an excellent instrument for measuring energy spectra with energies down to a few keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.