Abstract

Gas microstrip detectors have been integrated with low noise preamplifier and shaper electronics on a common silicon substrate. These devices were fabricated at the Hewlett-Packard company using a standard 0.8 μm process. This unique approach offers advantages over conventional microstrip design: ease of fabrication, lower noise, and higher channel density. The detector/electronics assembly was tested in a small drift chamber. An energy resolution of 18% FWHM and a noise level of 80 e − r.m.s per channel were obtained with a gas gain of about 200. These integrated microstrip detectors are being developed as the readout devices for a small, high-resolution Time Projection Chamber (MicroTPC). Low mass and high resolution make the MicroTPC well suited for use as a vertex detector, especially in high track-density environments such as RHIC and the LHC.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call