Abstract

Objective A fully automated method for reducing EOG artifacts is presented and validated. Methods The correction method is based on regression analysis and was applied to 18 recordings with 22 channels and approx. 6 min each. Two independent experts scored the original and corrected EEG in a blinded evaluation. Results The expert scorers identified in 5.9% of the raw data some EOG artifacts; 4.7% were corrected. After applying the EOG correction, the expert scorers identified in another 1.9% of the data some EOG artifacts, which were not recognized in the uncorrected data. Conclusions The advantage of a fully automated reduction of EOG artifacts justifies the small additional effort of the proposed method and is a viable option for reducing EOG artifacts. The method has been implemented for offline and online analysis and is available through BioSig, an open source software library for biomedical signal processing. Significance Visual identification and rejection of EOG-contaminated EEG segments can miss many EOG artifacts, and is therefore not sufficient for removing EOG artifacts. The proposed method was able to reduce EOG artifacts by 80%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.