Abstract

Testing of a newly developed position sensitive high count rate microstrip gas chamber (MSGC) detector at high count rate operation has been carried out at the Dutch–Belgian x-ray scattering beamline at the European Synchrotron Radiation Facility (Grenoble, France) with a high intensity x-ray beam. The measurements show local count rate capabilities up to ∼4.5×105 counts/s/channel. Experimental data taken with this detector are also shown. These tests show that both time resolution down to 1.5 ms/frame and a reliable operation at high counting rates can be achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.