Abstract
The sensitivity of the new generation of He-3 bolometers is such that it is possible to detect tiny atmospheric perturbations occurring at high altitude by means of a submillimetric photometer: an excess of 1% along one km of the stratosphere for H 2O vapour is detectable up to 10 km of altitude, while the same excess for O 2 is detectable up to 20 km and up to 35 km in the case of Ozone. In the present paper we describe a multi-band photometer operating at the focal plane of MITO (Millimeter and Infrared Testagrigia Observatory), capable of discriminating atmospheric perturbations in altitude and in chemical composition. The instrument is devoted to the study of the dynamics of the upper atmosphere in a long range program of far infrared observations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.