Abstract

A quality control programme for dosimetric aspects of mammography screening units is presented which consists of measurements at different phases of operation, i.e. acceptance, constancy and status tests. To investigate the adequacy of quality control procedures absorbed dose measurements were made at a mammography unit using an ionisation chamber and polymethylmethacrylate phantoms of various thicknesses. Major causes of instability of absorbed dose revealed by the quality control programme, were related to tube voltage, compensation of the automatic exposure control unit for radiographs of phantoms of varying thickness and film processing during periods of reduced frequency of operation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.