Abstract

The response of a neutron dosemeter based on plastic track detectors has been studied. The dosemeter (converter + detector) configurations used consist of a CR-39 detector 500 μm thick and a Makrofol converter 300 μm thick for fast neutrons (configuration 1) and 3 mm of air used as converter for thermal neutrons (configuration 2). The possibility of using Makrofol as a high energy neutron dosemeter has also been studied. In order to validate the results obtained from Monte Carlo simulations using the MCNP4A code, a set of irradiations by monoenergetic neutron beams has been performed at the Physikalisch-Technische Bundesanstalt (PTB), Gesellschaft fur Strahlen-Und Umweltforschung (GSF) and Paul Scherrer Institut (PSI) neutron irradiation facilities. Irradiations with realistic fields were performed at the Cadarache Nuclear Centre. An excellent agreement has been found between the simulated and the experimental values, not only for fast neutrons but also for thermal neutrons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.