Abstract

An integrating dosemeter system is described which terminates an X-ray treatment exposure after a pre-set dose of radiation, as measured by an ionization chamber placed between the tube and the patient. Details of the construction of an aluminium ionization chamber, which has proved more reliable than chambers made of Perspex, are given. The integrating process, including the discharge, is carried out electronically using a cold cathode trigger tube, and the mechanical counting is carried out by a chain of single digit counters which have previously been pre-set by means of electrical push-buttons to a figure calculated from the prescribed dose. The counters count down to zero and then terminate the exposure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.