Abstract

This paper presents the evaluation of the metrological performances of a DAQ-based sampling wattmeter (DAQ-SW) for the measurement of IEEE Std. 1459-2010 power quantities in non-sinusoidal conditions. The instrument makes use of two commercial data acquisition (DAQ) boards, a non-inductive current shunt, a personal computer and a commercial software for data processing. Starting from the results of an experimental characterization of the two instrument channels with non-sinusoidal voltages and currents and a correlation analysis, the combined standard uncertainties of IEEE Std. 1459-2010 quantities are evaluated, following the approach of the ISO Guide to the Expression of Uncertainty in Measurement (GUM).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call