Abstract
The performance of the next-generation BacT/ALERT® VIRTUO™ Microbial Detection System (VIRTUO™, bioMérieux Inc., Hazelwood, MO) was compared to the BacT/ALERT® 3D Microbial Detection System (3D, bioMérieux Inc., Durham, NC) using BacT/ALERT® FA Plus (FA Plus), BacT/ALERT® PF Plus (PF Plus), BacT/ALERT® FN Plus (FN Plus), BacT/ALERT® Standard Aerobic (SA), and BacT/ALERT® Standard Anaerobic (SN) blood culture bottles (bioMérieux Inc., Durham, NC). A seeded limit of detection (LoD) study was performed for each bottle type in both systems. The LoD studies demonstrated that both systems were capable of detecting organisms at nearly identical levels [<10 colony-forming units (CFU) per bottle], with no significant difference. Following LoD determination, a seeded study was performed to compare the time to detection (TTD) between the systems using a panel of clinically relevant microorganisms inoculated at or near the LoD with 0, 4, or 10 mL of healthy human blood. VIRTUO™ exhibited a faster TTD by an average of 3.5 h, as well as demonstrated a significantly improved detection rate of 99.9% compared to 98.8% with 3D (p-value <0.05).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European Journal of Clinical Microbiology & Infectious Diseases
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.