For γ-γ coincidence measurements a Compton-suppression spectrometer with a large solid angle of 120 msr and excellent suppression has been designed. The dimensions of the NaI anticoincidence shield have been optimized by means of Monte Carlo calculations. The NaI shield has a length of 35 cm and a diameter of 30 cm. A 20% efficient HPGe crystal with a dead layer of 0.22 mm and a hollow core is used as central detector. For 60Co, the average suppression of the Compton background between 100 and 1100 keV amounts to 11.8 with a photopeak loss of 5.8%. The areas of the 1173 and 1332 keV peaks taken together amount to 60% of the total number of registered counts for the suppressed spectrum and to 15% for the non-suppressed spectrum. The spectrometer has been tested in a γ-γ coincidence experiment with the reaction 24Mg + (45 MeV) 16O.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.