Abstract

The geometrical parameters which affect the suppression factor for a HpGe-NaI(Tl) Compton suppression spectrometer are studied by Monte Carlo simulation. The simulation parameters are designed for comparison of the experimental results of the specified Compton suppression spectrometer with a HpGe detector as the central detector. The simulated results are in coincidence with the experimental results. The optimal position of the HpGe detector within the shielding of the NaI(Tl) detector in two kinds of asymmetry Compton suppression spectrometer, which we named L(T)-shape and L(C)-shape is studied. The optimized suppression position is consistent with the previous simulated results, that the front face of the central detector is about at 3 4 – 4 5 length of the shielding along the hole dimension. Further simulation indicates that the suppression factor is quickly saturated by increasing the shielding thickness along the dimension perpendicular to the hole for the central detector.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.