Abstract

The newly introduced Nexfin(®) device allows analysis of the blood pressure trace produced by a non-invasive finger cuff. We compared the cardiac output derived from the Nexfin and PiCCO, using transcardiopulmonary thermodilution, during cardiac surgery. Forty patients with preserved left ventricular function undergoing elective coronary artery bypass graft surgery were studied after induction of general anaesthesia and until discharge to the intensive care unit. There was a significant correlation between Nexfin and PiCCO before (r(2) = 0.81, p < 0.001) and after (r(2) = 0.56, p < 0.001) cardiopulmonary bypass. Bland-Altman analysis demonstrated the mean bias of Nexfin to be -0.1 (95% limits of agreement -0.6 to +0.5, percentage error 23%) and -0.1 (-0.8 to +0.6, 26%) l.min(-1).m(-2), before and after cardiopulmonary bypass, respectively. After a passive leg-raise was performed, there was also good correlation between the two methods, both before (r(2) = 0.72, p < 0.001) and after (r(2) = 0.76, p < 0.001) cardiopulmonary bypass. We conclude that the Nexfin is a reliable method of measuring cardiac output during and after cardiac surgery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.