Abstract

Intrauterine pressure catheter (IUPC) is the primary device used to evaluate uterine activity. In contrast to the IUPC, electrical uterine myography (EUM) enables noninvasive measurement of frequency, intensity, and tone of contractions. The aim of this study was to determine the accuracy of EUM compared to IUPC. EUM measured myometrial electrical activity using a multichannel amplifier and a noninvasive position sensor. In all, 47 women in labor were monitored simultaneously with an IUPC and EUM. We compared the frequency, intensity, and tone of uterine contractions between the methods. The correlation of the frequency, intensity, and tone of contractions between uterine electromyography and IUPC was strong with significant r values of 0.808-1 (P < .0001). Electrical uterine electromyography yields information about uterine contractility comparable to that obtained with IUPC.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call