Abstract

The simplest stability indicating reversed phase Isocratic HPLC and UPLC methods has been developed and validated for the determination of fluconazole in bulk and solid pharmaceutical dosage form. A SunFire C18 (250 × 4.5 mm, 5 μm particle size) column has been used for HPLC and BEH C18 (100 × 2.1 mm, 1.7 μm particle size) column used for UPLC. The Mobile phase consisted of Methanol : Water (70 : 30) for HPLC and Methanol : Water (55 : 45 v/v) for UPLC. Isocratic flow was set at 1 mL/min and 0.30 mL/min, respectively, for HPLC and UPLC. For both HPLC and UPLC system detection has been performed at 211 nm with 30°C column oven temperature (good elution was obtained at 30°C) and injection volume, respectively, 2 μL and 20 μL for HPLC and UPLC.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.