Abstract

A composite cushion acetabular cup for a total hip replacement has been designed and developed jointly by Leeds University and DePuy International. In order to assess the long-term performance of this novel design, two sets of simulator tests of more than 4 million cycles duration have been carried out with the cushion bearings using the Leeds PA hip joint simulator with bovine serum as the lubricant. The results of these simulator tests were compared to the results from a previously reported study that used 32 mm ultrahigh molecular weight polyethylene (UHMWPE) acetabular cups. Under a physiological walking cycle simulation, with continuous cyclic motion and loading, the composite cushion cups produced negligible wear compared to a volumetric wear rate of 32 mm 3 per million cycles for the conventional UHMWPE acetabular cups. This study has demonstrated for the first time the beneficial effects of fluid film lubrication in reducing wear in composite cushion acetabular cups.