Abstract

Over the past 10 years, the number of dental implants has grown significantly. This increase has consequently led to an elevation of the statistics related to cases of peri-implantitis. Laser therapy has conquered a place among the therapies of excellence to treat peri-implantitis. However, the laser device used could influence the therapy’s success. The aim of this comparative experimental work was to highlight the differences in the work on grade 4 titanium surfaces of the most commonly used laser lights in this field, taking into consideration any structural damage that lasers could cause to implant surfaces. The lasers examined were a 980 nm diode laser; a 1064 nm Nd:YAG laser; and a new generation of 1064 nm Nd:YAG Q-switch nano pulsed laser. We evaluated the titanium temperature increase, the pre- and post-treatment two-dimensional surface appearance observed under the scanning electron microscope; finally, the three-dimensional pre- and post-treatment topographic analysis was assessed using atomic force microscopy. We showed that the 1064-nm Q-switch Nd:YAG nanosecond pulsed laser appears to be more suitable for the preservation of implant morphology because of the absence of the induction of metal damage.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.