Abstract
Using arbitrary Lagrangian-Eulerian (ALE) finite element method, this paper made a comparative study of the opening and closing behaviour of a downstream directional valve (DDM) and a St. Jude medical valve (SJM) through a two dimensional model of mechanical valve-blood interaction in which the valve is considered as a rigid body rotating around a fixed point, and the blood is simplified as viscous incompressible fluid. It's concluded that: (1) Compared with SJM valve, DDM valve opens faster and closes the more gently. (2) The peak back-flow of DDM is smaller than that of SJM. The present investigation shows that being a better analogue of natural valve, DDM has a brighter potential on its durability than SJM.