Abstract

An UV-Ion Mobility Spectrometer is a simple, rapid, inexpensive instrument widely used in environmental analysis among other fields. The advantageous features of its underlying technology can be of great help towards developing reliable, economical methods for determining gaseous compounds from gaseous, liquid and solid samples. Developing an effective method using UV-Ion Mobility Spectrometry (UV-IMS) to determine volatile analytes entails using appropriate gaseous standards for calibrating the spectrometer. In this work, two home-made sample introduction systems (SISs) and a commercial gas generator were used to obtain such gaseous standards. The first home-made SIS used was a static head­space to measure compounds present in liquid samples and the other home-made system was an exponential dilution set­up to measure compounds present in gaseous samples. Gaseous compounds generated by each method were determined on-line by UV-IMS. Target analytes chosen for this comparative study were ethanol, acetone, benzene, toluene, ethylbenzene and xylene isomers. The different alternatives were acceptable in terms of sensitivity, precision and selectivity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.