Abstract

A low voltage, low noise instrumentation amplifier (IA) applicable for Electrocardiogram signal acquisition system has been designed. The circuit is based on the current feedback topology which has been implemented using folded cascode structure at the input stage. It works at zero input common mode voltage with 1V supply. It has a wide input dynamic range of plusmn0.01 mV to plusmn1.2 mV. The voltage gain is 45 dB and it consumes 165 muW power. A high common mode rejection ratio (CMRR) of 125 dB has been achieved. The integrated input referred noise is 3 muVrms (0.2Hz - 150 Hz). The design has been done in 0.18 mum standard CMOS process.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.