Abstract
A new circuit topology for a low voltage, low noise instrumentation amplifier (IA) applicable for electrocardiogram signal acquisition system is designed. This circuit is based on the current feedback topology, which is implemented using folded cascode structure at both input and output stages. This helps to sense very low common mode voltage. Achieved input common mode range (ICMR) is 71.43 - 700 mV. DC electrode offset cancellation circuit is realized to remove DC electrode offset along with the low frequency noise. The IA can withstand a high DC electrode offset of ±18 mV. It also has a wide input dynamic range of ±1 ¿V to ±1 mV. The voltage gain is around 38 dB. It consumes 110 ¿W power with a supply voltage of 1 V. A high common mode rejection ratio (CMRR) of 80 dB is achieved. The integrated input referred noise is 1.64 ¿V <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">rms</sub> (0.1 Hz - 150 Hz). This design is implemented in 0.18 ¿m standard CMOS process.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.