Abstract
A current-mode interface chip for Silicon Photomultiplier (SiPM) array based positron emission tomography (PET) imaging front-ends is described. The circuit uses a high-speed current amplifier with a low input impedance, to minimize signal loss at the SiPM amplifier interface. To reduce the impact of dark noise, a novel high-speed threshold detection/comparator circuit is used to remove unwanted noise events. A prototype chip interfaces an array of SiPMs to the digital backend of a Positron Emission Tomography (PET) system using 64 readout channels, each of which contain a current amplifier and a threshold detection component. To reduce the number of backend channels, a row-column pulse positioning architecture (RCA) has been implemented. The ASIC occupies an area of 14.04 mm2 in 130nm STMicroelectronics HCMOS9GP process. The measured input impedance of the current amplifier is 20 ohms at 10 MHz, while the threshold detection circuit's propagation delay is 0.3-2ns.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Biomedical Circuits and Systems Conference : healthcare technology : [proceedings]. IEEE Biomedical Circuits and Systems Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.