Abstract

Sensitivity improvement has been achieved for a quadrupole mass spectrometer by replacing the flat repeller of a conventional ion source with a cylindrical repeller. Computer simulation of ion trajectories in the ion source demonstrated that ions were focused into the center of the ion lens with the cylindrical repeller but that ions were defocused with the flat repeller. Ion sources with the two repellers were tested for performance comparison using both a laboratory-built mass spectrometer and a commercial mass spectrometer. The cylindrical repeller showed about threefold sensitivity enhancement at lower repeller voltages across the mass range from 69to502m∕z.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.