Abstract

To assess dosimetric parameters in a case study where bilateral accelerated partial breast irradiation (APBI) is delivered using a strut-adjusted volume implant (SAVI) device. A 59-year-old female received APBI in both breasts over 5 days, with fractions of 3.4 Gy twice daily. A Vac-lok system was used for immobilization, and a C-arm was used for daily imaging. We generated dose-volume histograms (DVHs) for the brachytherapy plans to derive several important biologic factors. We calculated the normal tissue complication probability (NTCP), equivalent uniform dose (EUD), and tumor control probability (TCP) using the Lyman-Kutcher-Burman model parameters α = 0.3 Gy−1, α/β = 4 Gy, n = 0.1, and m = 0.3. In addition, we assessed the dose homogeneity index (DHI), overdose index, and dose nonuniformity ratio. D95 was >95% and V150 was <50 mL for both breasts. The DHIs were 0.469 and 0.512 for the left and right breasts, respectively. The EUDs (normalized to 3.4 Gy b.i.d.) were 33.53 and 29.10 Gy. The TCPs were estimated at 99.2% and 99.9%, whereas the NTCP values were 4.2% and 2.57%. In this clinical case, we were able to quantify the dosimetric parameters of an APBI treatment performed with a SAVI device.