Abstract

Several DP systems have been discussed and evaluated with respect to dynamic evacuation under actual operating conditions of an electron microscope. As a result, a cascade DP system has been found to be the most reasonable for an electron microscope where the column is evacuated by the first DP backed by the second DP, and for the camera chamber by the second DP with a large volume foreline backed by a mechanical pump. A spectrum of the residual gases in an electron microscope with a cascade DP system indicates a very clean vacuum with hydrocarbon partial pressures of less than 3×10−7 Pa.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.