Abstract

A novel bromide ion-selective PVC membrane sensor based on 2,3,10,11-tetraphenyl-1,4,9,12-tetraazacyclohexadeca-1,3,9,11-tetraene zinc(II)complex ( I) as carrier has been developed. The electrode exhibited wide working concentration range 2.2 × 10 −6 to 1.0 × 10 −1 M and a limit of detection as 1.4 × 10 −6 M with a Nernstian slope of 59.2 ± 0.5 mV per decade. The response time of electrode was 20 s over entire concentration range. The electrode possesses the advantages of low resistance, fast response and good selectivities for bromide over a variety of other anions and could be used in a pH range of 3.5–9.5. It was successfully used as an indicator electrode in the potentiometric titration of bromide ions with silver ion and also in the determination of bromide in real samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.