Abstract
A newly synthesized Schiff base 3-aminoacetophenonesemicarbazone (AAS) has been used for the preparation of Ni2+ selective PVC membrane electrode. The proposed electrode exhibits a Nernstian response over the nickel concentration range of 1.0×10−7 to 1.0×10−2molL−1 with a slope of 30.0±0.3mV/decade of concentration. The limit of detection as determined from the intersection linear segment of the calibration plot is 5.1×10−8molL−1. The electrode shows good selectivity towards nickel with respect to several alkali, alkaline earth, transition and heavy metal ions. The response time of the electrode is very fast (≥10s) and can be used for 17weeks in the pH range of 2.0–9.8. The electrode can also be used in partially non-aqueous media having up to 20% (v/v) methanol, ethanol or acetone content with no significant change in the value of slope or working concentration range. To investigate the analytical applicability of the electrode, it was successfully applied as an indicator electrode in Ni2+ ion potentiometric titration with EDTA, and in direct determination of nickel(II) in real samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.