A 20-channel electron cyclotron emission (ECE) detection system for the grating polychromator in JT-60U has been developed to measure electron temperature and its perturbations. Indium antimonide (InSb) detectors aligned in a row are cooled by liquid helium held by a 50 cm-diam cryostat, in which a hold time of the liquid helium reaches 7 weeks. Twenty-channel preamplifiers are dc-coupled ones, which enable to measure low-frequency ECE. From the comparison between the new system and the existing 20-channel detector system, where a refrigerator for cooling InSb detectors and ac-coupled preamplifiers are used, it is found that the signal-to-noise ratio is improved, and that small perturbations can be detected. By using this system, the electron temperature and its perturbations near the internal transport barrier in a reversed shear discharge are measured in detail.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.