Abstract

Low intensity transcranial focused ultrasound has been demonstrated to produce neuromodulation in both animals and humans. Primarily for technical reasons, frequency is one of the most poorly investigated critical wave parameters. We propose the use of a quadri-band transducer capable of operating at 200, 320, 850, and 1380 kHz for further investigation of the frequency dependence of neuromodulation efficacy while keeping the position of the transducer fixed with respect to the subject's head. This paper presents the results of the transducer calibration in water, in vitro transmission measurements through a monkey skull flap, 3-D simulations based on both a μ -computed tomography ( μ CT)-scan of a rat and on CT-scans of two macaques. A maximum peak pressure greater than 0.52 MPa is expected at each frequency in rat and macaque heads. According to the literature, our transducer can achieve neuromodulation in rodents and primates at each four frequencies. The impact of standing waves is shown to be most prominent at the lowest frequencies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.