Abstract
This work is focused on the testing of a new 68Ge/68Ga radionuclide generator based on cerium dioxide in polyacrylonitrile beads. During an 18-month period, parameters of elution were monitored, and basic radiolabelling studies were carried out. As current commercial solutions deal with high elution volume and low long-term stability, the constructed system offers several improvements. The composite sorbent provides uniformity of active component dispersion and high capacity for 68Ge. It enables minimal elution volume without fractionation (up to 1.8 mL) and stability of 68Ge breakthrough (under 0.001%) after initial period of wash-out and low cerium leakage (under 5 ppm).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.