Abstract

A functional clotting assay was recently reported to detect the factor V-Leiden mutation (R506Q) in patients receiving oral anticoagulants and in patients with Lupus Anticoagulant inhibitor. The original assay (Dahlback) to detect resistance to activated protein C (aPC-resistance) frequently gave unreliable findings in these patients. The change in the method is the use of bovine thromboplastin in a PT-derived assay, with a 1:10 sample dilution. In these conditions a reference normal plasma gives a prolongation of more than 35 seconds, while samples with a heterozygous or homozygous mutation give a prolongation respectively of 10–18 seconds or less 2 seconds. The test is easily automated and quickly performed on ACL/3000, and the reproducibility is good. © 1997 Elsevier Science Ltd

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.