Abstract
Gastrointestinal (GI) endoscopy is a common medical diagnostic procedure used for esophageal cancer detection. Current emerging capsule optoacoustic endoscopes, however, suffer from low pulse repetition rates and slow scanning units limit attainable imaging frame rates. Consequently, motion artifacts result in inaccurate spatial mapping and misinterpretation of data. To overcome these limitations, we report a 360º, 50 Hz frame rate, distal scanning capsule optoacoustic endoscope. The translational capability of the instrument for human GI tract imaging was characterized with an Archimedean spiral phantom consisting of twelve 100 µm sutures, a stainless steel mesh with a pitch of 3 mm and an ex vivo pig esophagus sample. We estimated an imaging penetration depth of ~0.84 mm in vivo by immersing the mesh phantom in intralipid solution to simulate light scattering in human esophageal tissue and validated our findings ex vivo using pig esophagus. This proof-of-concept study demonstrates the translational potential of the proposed video-rate endoscope for human GI tract imaging.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.