Abstract

A piezoelectric actuator for blood extraction pump in health monitoring system was developed. We devised the geometry of piezoelectric actuator with some slits, that allows the stretching and contracting deformation in in-plane direction to create large deflection in out-of-plane direction. The static behaviors under uniform electric field have been analyzed by finite element method. The computational results indicated that a bimorph actuator with cross slit under outside fixed condition is the most suitable for blood extraction pump. Abimorph actuator with cross slit l=4.0, w=0.5mm has been manufactured as an experiment. As a result, the developed actuator was proved to increase the deflection to 2.1 times at resonance frequency f=6460Hz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.