Abstract

To the Editor.— We read with interest the recent article by Kenney et al1 regarding the acceptability of the FIRST temp (Intelligent Medical Systems, Inc., Carlsbad, CA) tympanic membrane infrared temperature sensing device as a clinical tool for pediatricians.1 The authors conclude that the instrument provides body temperature measurements consistent with traditionally used methods (glass-mercury thermometer). We take exception with the analysis used to reach such a conclusion. The authors present their data as mean differences between FIRST temp and conventional methods with 95% confidence intervals around the mean.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.