Abstract
We compared the temperatures recorded, in febrile children admitted to a children's ward at Port Moresby General Hospital, by a doctor and by a group of nurses using glass mercury thermometers (GMT) and liquid crystal thermometers (LCT, Nextemp and Traxit. The mean difference (with 95% confidence intervals) in temperatures between GMT and Nextemp were -0.12 degrees C (-0.16 degrees C to -0.08 degrees C) for the doctor and 0.12 degrees C (0.04-0.20 degrees C) for nurses. The mean difference in temperatures between GMT and Traxit were -0.05 degrees C (-0.09 degrees C to -0.01 degrees C) for the doctor and 0.19 degrees C (0.10-0.28 degrees C) for the nurses. A similar result was obtained when one of the Nextemp thermometers used in the initial study was compared with GMT on a small sample of patients by the doctor 8 months later. Limited evaluation showed nursing staff were in favour of using the LCTs. Nextemp and Traxit thermometers can be used interchangeably with GMT in this setting.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.