Abstract
A synchronized accumulating radioisotope detector for radio gas chromatography was developed. It comprised seven gas-flow proportional counters each with an inner volume of 10 ml. Every counter tube was connected by a mutual anti-coincidence circuit to reduce the background. The transit time of gas particles in one counter tube could be set to an optimal value between 1 and 4 s by regulating the flow-rate of the counting gas, according to analytical requirements. The improved detector maintained high chromatographic resolution, which suggested the applicability of the apparatus to capillary gas chromatography.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.