Abstract
Individual isomers of octadecadienoic acid (C18:2) with conjugated double bonds (conjugated linoleic acids; CLA) exert different biological activities. Their distribution in food and tissues differs. Therefore, the separation of the various positional and geometric isomers is important. The time of analysis using silver ion-high performance liquid chromatography can extend up to 90 min. The aim of this study was to reduce this time. The time of analysis reduced from ca. 90 min onto 45 to 35 min, respectively, by the addition of 0.05% or 0.1% (v/v) 2-propanol to the mobile phase [acetonitrile (0.1%; v/v) and diethyl ether (0.5%; v/v) in n-hexane]. There was no effect on resolution of the 17 individual CLA isomers of the CLA mixture. Regarding the lowest coefficient of variation and an adequate baseline separation the use of 0.05% 2-propanol in the mobile phase is recommended, without any disadvantages and adverse effects on the service life of columns. In conclusion, adding 0.05% or 0.1% 2-propanol to the mobile phase shortens the time of analysis of CLA isomers, saves solvents and reduces costs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.