Abstract
Quadrupole Mass Spectrometers (QMS) is a very useful tool in vacuum process diagnosis. Tungsten filament based ion sources are vulnerable to contamination from process gas monitoring. Common symptoms of quadrupole mass spectrometer malfunction is appearance of unwanted contaminant mass peaks or no detection of any ion peaks. We disassembled used quadrupole mass spectrometer and found out black insulating deposits on inside of ion source parts. Five steps of cleaning procedure were applied and almost full restoration of functions were confirmed in two types of closed ion source quadrupole mass spectrometer. By using a numerical modeling (CFD-ACE+) technique, the electric potential profile of ion source with/without insulating deposit was calculated and showed the possibility of quadrupole mass spectrometer malfunction by the deterioration of designed potential profile inside the ion source.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Korean institute of surface engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.