This third lecture of “Basics of Vacuum measurement from viewpoint of Metrology” gives an example to estimate measurement uncertainties based on “Guide to the expression of Uncertainty in Measurement (GUM)” for calibrations of vacuum gauges. Both a capacitance diaphragm gauge (CDG) and a pirani gauge were calibrated by direct comparison to another CDG used as a reference gauge based on JIS Z 8750. Both the procedure to summarize the calibration results and the method to estimate the expanded uncertainty are explained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.