Abstract

Purpose. Noise occurs at measuring Electoretinogram(ERG) signals as the other bio-signal measurement. It is compared the denoising performance according to the mother function of wavelet transforms. Methods. The ERG signal that generated power supply noise and white noise was used as a sampling signal. The noise of ERG signal was filtered by using haar, db7, bior mother function. The filtering performance of each mother functions was compared using Fourier transform spectrum and SNR(signal to noise ratio). Results. In the haar functioin, the result of the Fourier transform spectrum was that the power supply noise is removed and the white noise performance is not good. The SNR was 27.0404. In the db7 function, the results of Fourier transform spectrum was that the power supply noise is removed and the white noise performance is good. The SNR was 35.1729. In the db7 function, the results of Fourier transform spectrum was that the power supply noise is removed and the white noise performance is the bset. The SNR was 35.4445. Conclusions. The db7, bior function was good results in power supply noise and white noise filtered. The bior function is suitable for filtering noise of the ERG signal.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.