Abstract

A fast response analyzer for measuring carbon dioxide concentration has been developed for transient characteristic and researches tested on internal combustion engine. The analyzer uses the well known NDIR(Non-Dispersive Infrared) method with miniaturized detection system, giving a time constant of approximately 30 microsecond, and sampling module consists of capillary tube. Since the transit time and the time constant of the sampling system depend on the sampling conditions, it is necessary to investigate the characteristics of sampling system before applied to exhaust gas measurement in engine. A unique method was designed to study the influence of the diameter of transfer sample line and operating conditions of the FRNDIR on transit time and time constant. A database of transit time and time constant was built up for different measured and simulated pressure conditions. The database can be used for correcting eventual concentration measurement.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call