Abstract

A simple and reproducible method for the rapid extraction and determination of trace amounts of cop- per(II) ions using mesoporous organo-silicas mesoporous silica and atomic absorption spectrometry is presented. Com- mon coexisting ions did not interfere with the separation and determination. The preconcentration factor was 100 (1 ml elution volume) for a 100 ml sample volume. The limit of detection of the proposed method is 1.0 ng ml . The maximum sorption capacity of sorbent under optimum conditions has been found to be 5mg of copper per gram of sorbent. The relative standard deviation under optimum conditions was 2.8% (n=10). Accuracy and application of the method was estimated by using test samples of natural and synthetic water spiked with different amounts of copper(II) ion.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.