Abstract
Thermal time constant measurement system was designed and fabricated to measure the thermal time constant, which shows a dynamic property of the thermocouple. Type K thermocouple samples were fabricated with variable shape each other and the thermal time constants of thermocouples were measured using the home-made thermal time constant measurement system. Thermal time constants of 12 type K thermocouple samples were distributed from 0.03 s to 8.2 s. It showed experimentally that the thermal time constant of thermocouple was increased linearly for the increase of the sheath diameter of thermocouple.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.